Read Type – continued
Continuous – The meter will continuously measure and update the display .
No indicator will be shown on the display . This read type is useful when
performing an experiment that requires continuous measurements to be
taken, regardless of the measurement stability . Press the log/print key
to export the measurement to the data log, if the data log is enabled in
the setup menu, and to a printer or computer, if a printer or computer is
connected to the meter and enabled in the setup menu .
Value Change In Measurement – The meter will display the measurement
when the reading reaches or exceeds the set high or low value that is
programmed by the operator . When the reading reaches or exceeds the set
high or low value, the meter will automatically export the measurement to
the data log, if the data log is enabled in the setup menu, and to a printer or
computer, if a printer or computer is connected to the meter and enabled in
the setup menu . If Value Change In Measurement is selected, press the ▲ / ▼
keys to highlight Min Value, enter the minimum measurement value using
the numeric keypad and the decimal key, press the ▲ / ▼ keys to highlight
Max Value, enter the maximum measurement value using the numeric
keypad and the decimal key and press the f3 (next) key .
Timed Reading – The meter will display a measurement after a set
time delay and lock and hold the measurement after the time delay is
reached . To start another time delay cycle, press the measure key .
Once the measurement is locked, the meter will automatically export the
measurement to the data log, if the data log is enabled in the setup menu,
and to a printer or computer, if a printer or computer is connected to the
meter and enabled in the setup menu . If Timed Reading is selected, enter the
time delay value in a minutes:seconds format using the numeric keypad and
press the f2 (accept) key .