Calibration Setup – continued
ISE calibration settings – Entering the Calibration Setup menu will access
the Autoblank, Low Level Stability and Isopotential settings .
Autoblank – Press the ▲ / ▼ keys to highlight Autoblank and press the
f2 (select) key . Press the ▲ / ▼ keys to highlight Off or On and press
the f2 (accept) key .
Low Level Stability – Press the ▲ / ▼ keys to highlight Low Level Stability
and press the f2 (select) key . Press the ▲ / ▼ keys to highlight Off or
On and press the f2 (accept) key .
Isopotential – Press the ▲ / ▼ keys to highlight Isopotential and press
the f2 (select) key .
• To turn off or on temperature compensation for ISE measurements,
press the ▲ / ▼ keys to highlight ISO On/Off and press the f2 (select)
key . Press the ▲ / ▼ keys to highlight Off or On and press the
f2 (accept) key .
• To enter the ISE isopotential point, press the ▲ / ▼ keys to highlight
ISO Value and press the f2 (select) key . Use the numeric keypad and
decimal key to enter the value and press the f2 (accept) key .
• Once all of the ISE isopotential settings are entered, press the
f1 (back) key .
Once all of the ISE calibration settings are entered, press the f1 (back) key .