Press the ▲ / ▼ keys to highlight Auto-Read, On Ready, At Time Intervals,
Continuous, Value Change In Measurement or Timed Reading as the read type
and press the f2 (accept) key to save the setting .
i . Auto-Read – The meter will display the measurement as it stabilizes and
lock and hold the measurement when it is stable . Once a measurement
is locked, press the measure key to take a new measurement .
Once the measurement is locked, the meter will automatically export
the measurement to the data log, if the data log is enabled in the
setup menu, and to a printer or computer, if a printer or computer is
connected to the meter and enabled in the setup menu .
ii . On Ready – The meter will show stabilizing on the display until the
measurement is stable . When the measurement is stable, the meter
will show ready on the display . The display will automatically update if
the measurement changes and the measurement will not lock . Once
ready is shown on the display, the meter will automatically export
the measurement to the data log, if the data log is enabled in the
setup menu, and to a printer or computer, if a printer or computer is
connected to the meter and enabled in the setup menu .
iii . At Time Intervals – The meter will read and display a measurement at
the set time interval that is programmed by the operator . At the set
time interval, the meter will automatically export the measurement
to the data log, if the data log is enabled in the setup menu, and to a
printer or computer, if a printer or computer is connected to the meter
and enabled in the setup menu . If At Time Intervals is selected, enter
the time interval value in a hours:minutes:seconds format using the
numeric keypad and press the f2 (accept) key .
iv . Continuous – The meter will continuously measure and update the
display . No indicator will be shown on the display . This read type is
useful when performing an experiment that requires continuous
measurements to be taken, regardless of the measurement stability .
Press the log/print key to export the measurement to the data log, if
the data log is enabled in the setup menu, and to a printer or computer,
if a printer or computer is connected to the meter and enabled in the
setup menu .