Technical Data Pa 17 - Siemens 7MC3040-... Operating Manual

Digital radiation pyrometer
Table of Contents

Advertisement

s

21 Technical Data PA 17

Measuring ranges :
(adjustable in partial
ranges):
400 ... 2000 °C
Sensor:
Thermopile
Spectral sensitivity:
CO
band
2
Focussing range:
0.8 m ... 
Distance to target size
ratio:
75:1 at 800 mm
Digital output:
Periodic output of measure-
ment data with adjustable
cycle time
Analogue output 1 & 2:
0(4) ... 20 mA linear,
switchable, scalable
(4...20 mA normally)
Resistance:
max. 500 
Reponse time t 98 :
 100 ms
Resolution Analogue output:
0.2 K + 0.03 % of the adjusted
span
Resolution Display:
1 K
Resolution USB / RS 485:
0.1 K at terminnal operation
Measuring uncertainty:
0.75 % of reading +1 K
(at ε =1.0 and TA=23 °C)
Linearisation:
digital via microcontroller
Repeatability:
2 K
Sighting device:
through-the-lens sighting with
target marking, laser spot light
or integrated camera
Ambient operating
temperature:
0 ... 65 °C
Indication of Overheating:
If the internal temperature of the
pyrometer exceeds 80 °C the
analog output will show a value >
20.5 mA
Storage temperature:
-20 ... 80 °C
Permissible humidity:
95% r.H. max.
(non-condensing)
Temperature coefficient with
reference to 23 °C:
≤ 0.04 %/K
Interface:
USB / RS485 with integrated
software to set parameters and
transmit measurement data to a
PC
Analogue input:
0 - 10 V
Digital output:
2 Open collector outputs
24 V; ≤ 30 mA
Digital input:
2 to 24 V
Power supply requirements:
24 V DC +10% / -20%
current input ≤135 mA / ≤175
mA with camera, 150 mA with
switched on spotlight
Ripple: ≤ 200 mV
Dimensions:
Ø 65 x 220 mm
(when connector is attached)
Housing material:
Stainless steel
64
Operating manual PA
Mounting:
External thread M 65 x 2
length 40 mm
Weight:
Approx. 0.9 kg
Connection:
with 8-pin connector
Protection:
IP 65 according to
DIN 40050
(when connector is attached)
Adjustable parameters:
Analogue output 1 & 2:
source/ scaling
Digtal input output 1 & 2:
source/ switch point
Transmission factor
Compensation of background
radiation
Look-up table for temperature
alignment
Emissivity ε:
10 - 110 %
increment size 0.1 %
Smoothing function t
98
0 – 999 sec
Memory modes:
- Min./Max. (peak picker)
- Double maximum with
adjustable hold time
Optional accessories:
calibration certificate
according to ISO 9001
calibration certificate
according to DKD
large variety of mounting
devices, digital displays,
software, etc.
:

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents