The cycle time in which the temperature reading is transmitted can be
set at the PC terminal (minimum cycle duration is 0.1 second).
12.2

Terminal connection via USB

For communication through a terminal connection via USB set the pa-
TERM
rameter
TERM
The parameter
TERM
Parameter
 Install the pyrometer's USB driver on the PC
 Connect the pyrometer with a USB cable to the PC
 Start a standard terminal program (e.g. Windows Hyperterminal or
Putty)
 Select the correct COM port
 Set the interface parameters for the serial interface (see chapter
transmission of measurement values)
 Open the connection
12.3

Terminal connection via RS485

For communication through a terminal connection via RS485 set the pa-
TERM
rameter
TERM
The parameter
TERM
Parameter
Caution !
When using a terminal connection via the RS485 interface, Cel-
laView cannot be used via the RS485 interface any longer.
A USB/RS485 converter is needed when working with the RS485 inter-
face. The distance between pyrometer and PC can be up to 1200 m.
In order to prevent reflections on the RS485 connection, it is absolutely
necessary to terminate the connection at the pyrometer and at the con-
verter. The termination is integrated in the pyrometer and is active when
delivered from the factory.
on the pyrometer to USB (default setting).
is available on configuration layer C011.
USB
. =
on the pyrometer to
is available on configuration layer C011.
. =
R485
Operating manual PA
R485
.
45

Hide quick links: