Siemens 7MC3040-... Operating Manual page 57

Digital radiation pyrometer
Table of Contents

Advertisement

Operating manual PA
s
applies to data acquisition parameters and input/output settings. Use
keys "B", "C" and "D" for direct access to enable the adjustments.
If you make a mistake while making the adjustments, simply enter off-
set=0.0 und factor=1.0, or set User Cal. to „off".
Command "A" redefines the pyrometer's entire measuring range. This
new temperature span may be smaller or larger than the range originally
programmed by the manufacturer. When selecting a new temperature
range for your ARDOCELL PA, make absolutely sure that the tempera-
ture span you select is actually covered by the pyrometer model you
have purchased. The only way to be certain is to perform measurement
tests.
Press "S" to enter a short text to name the control point. View this text by
selecting „Q" in the Main Menu.
49