User-defined calibration / scaling of the current out-
put
If necessary, the pyrometer can be adjusted with a user-defined calibra-
tion function. The following drawing explains the effects for offset and
factor.
T
shall
Caution:
To recalibrate your ARDOCELL PA, you will require a calibration
furnace and a reference standard.
In addition to a user-defined calibration, the user-defined range function
can also be activated. With this function being active, scaling of the cur-
rent output can be extended.
Please note:
The measuring accuracy is not defined when the scal-
ing is outside the pyrometer range.
Operating manual PA
Current state
Influence - factor
Influence - offset
T
Pyrometer
47