Section 4: Controlling Multiple Series 2600s (TSP-Link)
node[4].smua.reset()
node[1].smua.reset()
Using the alias (localnode)
The variable localnode is an alias for node[N] , where N is the node number of the Master.
For example, assume that Node 1 is the Master in a TSP-Link system. To send a command to
Node 1, you can use node[1] or localnode as follows:
node[1].smua.reset()
localnode.smua.reset()
Running scripts in a TSP-Link system
For remote programming, only a script stored in the Master can be run. Scripts in the Slaves
cannot be accessed. See
For front panel operation, a script in any node can be run from its front panel. For details, see
do I run a script from the front
4-4
Series 2600 System SourceMeter
-- Resets SMU A of Node 4.
-- Resets SMU A of Node 1.
-- Resets SMU A of the Master (Node 1).
-- Resets SMU A of the Master (Node 1).
"How do I interact with scripts using Test Script
panel?" on
page
3-2.
Return to
Section Topics
®
Instruments User's Manual
Builder?" on
page
3-3.
"How
2600S-900-01 Rev. C / January 2008