Keithley 2601 User Manual page 103

Series 2600 system sourcemeter instruments
Hide thumbs Also See for 2601:
Table of Contents

Advertisement

Series 2600 System SourceMeter
Figure B-3
Stacking channels for higher voltage
NOTE Each stacked channel adds approximately 100μA of common mode
2600S-900-01 Rev. C / January 2008
®
Instruments User's Manual
current that is seen by the channels below it. Therefore, in the
example above, Model 2602-1 Chan A will measure approximately
100μA x 3 = 300μA higher than Model 2602-2 Chan B. Similarly,
Model 2602-1 Chan B will measure approximately 100μA x 2 =
200μA higher than Model 2602-2 Chan B. The most accurate
measurement of actual current flowing through the DUT will be
obtained from Model 2602-2 Chan B.
Return to
Section Topics
Appendix B: Frequently Asked Questions
B-7

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

26022612261126352636

Table of Contents