Series 2600 System SourceMeter
The above command defines a display input field and message prompt for entering the bias
voltage. The -40 parameter sets the minimum voltage that can be entered, and the 40 parameter
sets the maximum voltage that can be entered.
Now assume you wish to change the bias input limits to ±10V. Edit the line of code to the following:
92bias = display.prompt("+00.000", " Seconds", "Enter BIAS Voltage.", 0, -
Saving the script – After modifying the script file, it must be saved by clicking the save icon on the
toolbar of the Test Script Builder as follows:
Figure 3-3
KIGeneral project imported into the Test Script Builder
Step 3: Storing the script in non-volatile memory of the Series 2600
Reference
After modifying a factory script, it can be loaded back into the Series 2600 and saved in non-
volatile memory as a user script.
2600S-900-01 Rev. C / January 2008
®
Instruments User's Manual
10, 10)
See "User scripts" in Section 2 of the Series 2600 Reference Manual for more
information on storing user scripts in non-volatile memory.
Return to
Section Topics
Section 3: Test Script Processor Interaction
3-7