Section 3: Test Script Processor Interaction
Figure 3-5
Run configuration example - Script Attributes tab shown
Running the user script
Reference
Remote programming
The functions for the KIGeneralUser user script can be called after the script is run as follows:
KIGeneralUser()
After the above command is executed, any of the functions that make up the user script can be
called in the same manner that factory script functions are called (see
page
3-3).
Front panel operation
To run the script from the front panel, a test name has to be added to the USER TESTS submenu
of the LOAD TEST menu. Since the KIGeneralUser script is made up of eight functions, a front
panel user test must run the script and also call a function. The following command will add the
test name "PulseVMeasI" to the menu. It will also configure the test to run the KIGeneralUser
script and call the PulseVMeasureI function:
display.loadmenu.add("PulseVMeasI", "KIGeneralUser() PulseVMeasureI()",
display.SAVE)
3-10
See "User scripts" in Section 2 of the Series 2600 Reference Manual for details on
running a user script.
Return to
Section Topics
Series 2600 System SourceMeter
"Step 1: Call the
2600S-900-01 Rev. C / January 2008
®
Instruments User's Manual
function" on