FAN   R EPLACEMENT  
 
 
 
When   r eplacing   t he   f an,   a lways   u se   a n   E SD   s afe-­‐work   a rea.   A lso,  
use   a n   E SD   w rist   s trap   t o   p rotect   e lectronic   c ircuitry.    
 
 
1.   Remove   t he   P SA   s ystem   p er   t he   P SA   R EMOVAL   s ection   o f   t his   m anual.    
 
2.   Disconnect   t he   M olex   c onnector   f rom   t he   c ircuit   b oard   t o   t he   f an.  
 
 
3.   Remove   t he   f our   P hilips   h ead   s crews   t hat   s ecure   t he   f an   t o   t he   c enter   s ection.  
 
 
Portable   O xygen   C oncentrator    
Figure   2 4.1:   F an   M olex   C onnection  
Figure   2 4.2:   F an   S crew   R emoval  
 
 
 