FEED/WASTE   M ANIFOLD   R EPLACEMENT  
 
1.   Remove   t he   P SA   s ystem   a ssembly   p er   t he   P SA   R EMOVAL   s ection   i n   t his   m anual.  
 
2.   Carefully   r emove   t he   t ie   w raps   t hat   b undle   t he   s olenoid   v alve   w ires   t ogether.  
 
3.   Loosen   a nd   r emove   t he   c enter   b olt   f rom   t he   P SA   a ssembly   u sing   a   5 /32"   A llen   w rench.  
 
4.   Holding   b oth   s ieve   b eds,   g ently   p ull   u pwards,   r emoving   t he   p roduct   m anifold   a nd   s ieve  
beds.   S et   t he   p roduct   m anifold   a nd   s ieve   b ed   a ssembly   a side,   k eeping   t he   t op   o f   t he   s ieve  
beds   c onnected   t o   t he   p roduct   m anifold.  
 
5.   Remove   t he   f eed/waste   m anifold   b y   p ulling   u pward   o n   t he   m anifold.   B oth   f ittings   f rom   t he  
manifold   t o   t he   c ompressor   s hould   s till   b e   c onnected   t o   t he   m anifold.   I f   n ot,   r emove   t he  
fitting(s)   f rom   t he   c ompressor.  
 
 
 
 
 
 
 
 
 
 
                                                        F igure   1 7.1:   R emoving   t he   F eed/Waste   M anifold  
Portable   O xygen   C oncentrator    
 
 
 
MN220-­‐1   A             6 9  