Solenoid   V Alve   R Eplacement - CAIRE FreeStyle Provider Technical Service Manual

With ultrasense and patient-removable battery
Hide thumbs Also See for FreeStyle:
Table of Contents

Advertisement

 
 
3.   Remove   t he   t ie   w raps   t hat   b undle   t he   s olenoid   v alve   w ires   t ogether.  
 
4.   Loosen   a nd   r emove   t he   c enter   b olt   f rom   t he   P SA   a ssembly   u sing   a   5 /32"   A llen   w rench.  
 
5.   Holding   t he   s ieve   b eds   i n   p lace,   g ently   p ull   u pward   o n   t he   p roduct   m anifold   t o   r emove.  
Keep   t ubing   a ssembly   i n   p lace.    
 
6.   Align   t he   n ew   p roduct   m anifold   w ith   t he   o utlet   o f   t he   s ieve   b eds,   a nd   g ently   p ush   t he  
manifold   d own   i nto   p lace.  
 
7.   Reconnect   t he   t ubing   a ssembly   t o   t he   n ew   p roduct   m anifold.  
 
8.   Place   c enter   b olt   t hrough   t he   p roduct   m anifold   a nd   t ighten   t o   1 5   i n-­‐lbs.    
 
 
SOLENOID   V ALVE   R EPLACEMENT  
 
 
 
Turn   O FF   t he   u nit,   r emove   t he   b attery,   a nd   d isconnect   t he   p ower  
cord   f rom   t he   e lectrical   o utlet   b efore   s ervicing   t o   p revent  
accidental   e lectrical   s hock   a nd   b urn   h azard.    
 
Portable   O xygen   C oncentrator    
 
Figure   1 5.4:   C enter   B olt   R emoval  
 
 
 
MN220-­‐1   A             6 5  