the   h istorical   h our   m eter   w ill   f unction.   T he   h our-­‐glass   i con   w ill   f lash   t o   i ndicate   t hat   t ime   i s   b eing  
accumulated.    
 
Toggle/Reset   S witch  
This   s witch   ( showing   i n   F igure   2 0)   t oggles   t he   d isplay   o r   r esets   t he   m aintenance   c hannel  
depending   o n   t he   a ctivation   t ime   a nd   t he   c urrent   d isplay   m ode.  
 
Upon   p ower-­‐up,   t he   d isplay   d efaults   t o   t he   h istorical   d isplay   m ode.   T he   d isplay   w ill   r emain   i n  
historical   m ode   u ntil   t he   p ushbutton   i s   u sed   t o   t oggle   t he   d isplay   i nto   m aintenance   m ode.   T he  
hour   m eter   w ill   r emain   i n   t his   m ode   u ntil   t oggled   o r   u ntil   p ower   i s   r emoved.   T he   h our-­‐glass  
icon   w ill   r emain   o n   a nd   t he   " TMR1"   i con   w ill   b e   d isplayed   b elow   t he   h our-­‐glass   i con.   A s   w ith  
the   h istorical   d isplay   m ode,   t he   h our-­‐glass   i con   w ill   f lash   t o   i ndicate   t hat   t ime   i s   b eing  
accumulated   a nd   t he   T MR1   i con   w ill   r emain   o n   s olid.   T he   h istorical   c hannel   c annot   b e   r eset.   T o  
reset   t he   m aintenance   c hannel,   b egin   w ith   t he   m odule   i n   m aintenance   m ode,   a nd   t hen   p ress  
the   t oggle/reset   s witch   f or   f ive   s econds.   A   r eset   i s   i ndicated   b y   a   " 0"   d isplay.    
 
HOUR   M ETER   R EPLACEMENT   F OR   F REESTYLE   O NLY  
 
 
When   r emoving   t he   f ront   c over   t o   r eplace   t he   h our   m eter   o n   a  
FreeStyle,   a lways   u se   a n   E SD   s afe-­‐work   a rea.   A lso,   u se   a n   E SD   w rist  
strap   t o   p rotect   e lectronic   c ircuitry.    
 
 
1.   Remove   t he   f ront   c over   h ousing   p er   t he   H OUSING   R EMOVAL   s ection   o f   t his   m anual.  
 
2.   Ensure   t hat   a ll   p ower   s ources   a re   d isconnected.    
 
3.   Locate   t he   h our   m eter   ( shown   i n   F igure   2 0).  
 
4.   Using   a   s mall   P hilips   h ead   s crew   d river,   r emove   t he   2   s crews.  
 
5.   Gently   p ull   u pward   t o   r emove   t he   h our   m eter   f rom   t he   8 -­‐pin   c onnector   ( J3).   W hen  
removing,   t ake   c are   n ot   t o   b end   o r   b reak   a ny   p ins   o n   t he   J 3   c onnection   h eader.  
 
6.   Replace   h our   m eter.    
 
 
FREESTYLE   5   H OUR   M ETER   F UNCTIONALITY  
The   F reestyle   5   l ogs   t he   h ours   o f   u se   o n   t he   i nternal   c ircuit   b oard.   A n   i nterface   k it   c an   b e   p urchased  
that   a llows   y ou   t o   r etrieve   t his   h our   r eading.   T he   H our   M eter   K it   c an   b e   o rdered   t hrough   C AIRE  
Customer   S ervice   ( 1-­‐800-­‐482-­‐2473)   b y   u sing   P N   K I406-­‐1.  
 