Do   n ot   u se   l iquid   d irectly   o n   t he   F reeStyle   o r   F reeStyle   5   u nit   t o   c lean   i t.   A   l ist   o f  
undesirable   c hemical   a gents   i ncludes   b ut   i s   n ot   l imited   t o   t he   f ollowing,  
according   t o   t he   p lastics   m anufacturer:   a lcohol   a nd   a lcohol-­‐based   p roducts,  
 
concentrated   c hlorine-­‐based   p roducts   ( ethylene   c hloride),   a nd   o il-­‐based  
products   ( Pine-­‐Sol,   L estoil).   T hese   a re   N OT   t o   b e   u sed   t o   c lean   t he   p lastic  
housing   o n   t he   F reeStyle   o r   F reeStyle   5 ,   t hey   c an   d amage   t he   u nits'   p lastic.  
 
 
Clean   o r   r eplace   t he   a ir   i ntake   /   g ross   p article   f ilter   w ith   w arm   s oapy   w ater   b etween   e ach  
patients   u se.   T he   p atients   s hould   b e   i nstructed   t o   c lean   t his   f ilter   a t   l east   o nce   p er   w eek,  
depending   o n   t he   e nvironment,   d uring   n ormal   o peration.  
 
Allow   t he   u nit   t o   a ir   d ry.   T hen   r e-­‐test   o xygen   c oncentration,   t est   t he   d uration   o f   t he   b attery  
pack,   a nd   e nsure   t he   A C   a nd   D C   p ower   s upplies   a re   i n   g ood   w orking   c ondition   b efore   y ou  
return   t he   u nit   t o   y our   i nventory.  
 
Make   t he   b leach   s olution   a   1 :100   d ilution   o f   5 .25%   s odium   h ypochlorite.   M ix   o ne  
part   h ousehold   b leach   ( e.g.,   C lorox   w ith   9 9   p arts   c old   t ap   w ater).   T o   m easure   t he  
solution   e asily,   t ake   1 /4   c up   o f   h ousehold   b leach,   a nd   m ix   i t   w ith   a   g allon   o f   c old  
tap   w ater.   A llow   t he   m ixture   t o   s it   o n   p otentially   c ontaminated   s urfaces   f or   t en  
 
minutes.  
 
 
CAIRE   d oes   n ot   r ecommend   t he   s terilization   o f   t his   e quipment.  
 
 
RECORDING   M AINTENANCE    
Whenever   m aintenance   o r   s ervice   i s   p erformed   o n   a   u nit,   a n   e ntry   s hould   b e   m ade   i n   t he  
service   l og   f or   t hat   c oncentrator   o r   r ecorded   i n   a ccordance   w ith   y our   c ompany's   s tandard  
procedure.  
 
 
 
 
 
 
 
 
 