4.   Before   r eplacing   t he   b oard,   e nsure   t hat   t he   m ain   c ircuit   b oard   a nd   a ll   p ins   a re   s traight   a nd  
unbroken   ( intact).  
 
5.   Locate   t he   c omposite   s ide   m arked   " top",   o r   t he   s ide   w ith   t he   s ingle   r ow   c onnector,   a nd  
properly   a lign   t he   c onnections.    
 
 
 
6.   View   t he   r eplacement   b oard   f rom   t he   s ide   a nd   a t   a n   a ngle   t o   m ake   c ertain   t hat   t he   p ins   a re  
aligned   p roperly   b efore   s eating   t he   c onnector.  
 
7.   When   t he   r eplacement   b oard   i s   p roperly   a ligned,   g ently   p ress   t he   b oard   i nto   p lace   u ntil  
properly   s eated.    
 
 
Portable   O xygen   C oncentrator    
Figure   2 1.1:   R emoving   M otor   C ontroller   B oard  
Figure   2 1.2:   A ligning   P ins   t o   t he   C onnection  
 
 
 
 