Adjustments
5·15.
OHMS ADJUSTMENT.
5-16. Equipment Required.
Standard Resistors:
±
K
ohm
.0005070
9330A/l K)
10
K
ohm
.001 %
±
9330AllOK)
100 K ohm
.001 % (Guildline 9330/ 100K or
±
9330A/1OOK)
1 M ohm
.002% (Guild line 9330/1M)
±
10 M ohm
.01 % (Guild line 95206)
±
5-17. Adjustment Procedure.
a. Press the 3456A's RESET button and set the in
strument to the 6 Digit Display and 4-Wire Ohms con
figuration.
b. Connect the 10 K ohm Standard Resistor to the
3456A 's input terminals. Adjust calibration pot "F" for
± I
a 10.00000
3
count reading. Disconnect the Stan
+
dard Resistor.
NOTE
The Standard Resistors are connected in the
4-
Wire con f iguration, as shown in Figure
5-4.
RATlO REF
4'o1R
n SENSE
STANDARD
RESISTOR
Figure
5·4.
Standard Rasistor Connection.
c. Connect the 1 K ohm Standard Resistor to the in
put
terminals.
Adjust calibration pot "G"
±
1000.000
1 count reading. Disconnect the Standard
Resistor.
d. Connect the 100 K ohm Standard Resistor
input terminals. Adjust calibration pot "H"
+
100.0000
3
1 count reading. Disconnect the Stan
±
dard Resistor.
e. Connect the 1 M ohm Standard Resistor to the in
put terminals. Adjust calibration pot
-3
1 count reading. Disconnect the Standard Resistor.
±
5·4
(Guildline
9330/ l K
or
(Guildline
9330/ l O K
or
VOl IS
"AA
"AA
GUARD
IQ]
1056·5-4
for a
the
10
for a
" I"
for a 1000.000
f. Connect the 10 M ohms Standard Resistor to the
input terminals . Adjust calibration pot
±
10.000<XJ
6
4 counts reading. Disconnect the Stan
+
dard Resistor. This completes the Ohms Adjustments.
5·18.
ACV ADJUSTMENTS.
5-19. Equipment Required.
AC Calibrator (Fluke Model 5200A)
5-20. Adjustment Procedure.
a. Press the 3456A's RESET bunon and set the in
strument to the ACV fuction.
b. Set the AC Calibrator for a
connect it to the 3456A's VOLTS input terminals.
c. Set the 3456A to the 100 V range and adjust the in
strument's calibration pot "K" for a 0 1 . 000
reading.
d. Set the 3456A to the
tion pot
"L"
for a 1 .00000
e. Set the 3456A to the 1000 V range and adjust either
"N"
calibration pot "M" or
reading.
f.
Repeat Steps c, d, and e in the given order until all
the readings are within the limits.
NOTE
"K", "L",
Calibration pots
interact with each other and should be ad
justed until the readings converge.
g. Disconnect the AC Calibrator from the 3456A.
Short the 3456A's input terminals.
h. Set the 3456A to the ACV
V range. Adjust calibration pot
reading.
i. Set the 3456A to the ACV function and do the
following:
1 . Note the 3456A's displayed reading.
2. Adjust calibration pal
reading.
3. Continue adjusting calibration pot "M" in the
same direction until the reading noted in Step 1 is
reached (within
2 counts).
±
j . Repeat Steps h and i until both readings converge
±
with each other (within
Model 3456A
"J"
for a
I
V,
I
kHz output and
± I
count
I
V range and adjust calibra
±
3 counts reading.
±
for a 001.00
2 counts
N"
and " M or
+
DCV function and the
"N"
for a minimum
"M" for a minimum
2 counts).
)