If You Have a Problem
Other Problems

If the test results are not what you expected

If the test results are not what you expected, it could be because of one of the
following:
Do not exceed the specified power input limit for the analyzer.
C A U T I O N
•I The input signal was not connected.
q
q
4 - 1 4
Ensure that the signal is present at the analyzer input.
l
Ensure that the coupler connections and rear-panel connections are
l
made. See Chapter 2 for more information.
Ensure proper test signals, line location and field number were specified
l
for the Option 107 required tests.
The input signal is too low to be measured.
If the input signal is too low, use a preamplifier to boost the input signal
l
(ensure that the voltage and power of the input signal does not exceed
the specified input limits for the analyzer). See "Cable TV Measurement
Specifications and Characteristics" in Chapter 9 for more information
about the minimum carrier level needed.
The wrong tune configuration was selected.
Ensure that the correct tune configuration has been selected.
l
If you are not using an external pad, ensure that EXT PAD YES NO
l
system" in Chapter 3 for more information.