Illegal Command; Invalid Symtab Entry: Symtab Overflow - HP 8591C User Manual

Cable tv analyzer cable tv measurements
Table of Contents

Advertisement

If You Have a Problem
Error Messages
If you see this message while loading the HP 85721A personality into analyzer
memory, it indicates that the analyzer's firmware must be updated before the
personality can be used.
Check the Ermware date for your analyzer:
q
Turn on the analyzer. The Ermware date is displayed on the analyzer's
l
display when the analyzer is Erst turned on. The date is displayed in a
year, month, day, format. For example, if rev 93.09.23 is displayed,
the release date of the Ermware was September 23, 1993.
To solve this problem:
If the release date for the analyzer's firmware was prior to September 23,
l
1993, the firmware must be updated to use the HP 85721A personality.
Contact your HP sales office for more information about updating analyzer
Ermware.
This error message indicates that there was not enough available analyzer
memory.
To solve this problem, you must delete the other programs in the analyzer
memory as follows:
1. Press @YZZZYJ
ERASE DLP MEM , (W).
3. Reload the HP 85721A personality using the procedure "Step 1. Load the
HP 85721A personality into analyzer memory" in Chapter 1.
4-6