Description - HP 8591C User Manual

Cable tv analyzer cable tv measurements
Table of Contents

Advertisement

The default value for the channel tuning configuration is the standard (STD)
channel format. In addition, the USR CN 0 CNTR FRQ and USR CH 0 SPAN
channel. This channel is then accessed by selecting channel 0 (zero).
The default settings are for channel 2 (STD). These settings are stored in
non-volatile analyzer memory. This means that the analyzer retains the
selections, even when power is turned off, until a new selection is made.
Press CABLE TV ANALYZER , Setup , to access Channel Tuning
See Also
"Step 1. Configure the test system" in Chapter 2.
Allows you to specify the location above or below a visual carrier to measure
system CSO and system CTB. The defaults are channel 4, CTB 6 MHz above
the visual carrier, and CSO 4.75 MHz above the visual carrier.
Simply press CSO FREQ OFFSET (CSO) and/or C'I'B FREQ DFFSET (CTB),
enter a positive frequency to measure above the visual carrier or a negative
frequency to measure below the visual carrier, then press CENTER_).
Press CABLE TV ANALYZER, Setup , System Beats to access
CSO FREQ OFFSET and CTB FREQ OFFSET.
See Also
"Step 1. Configure the test system" in Chapter 3.
TV Receiver, Setup, and AUX CTRL Menu Maps and Softkey Descriptions
7 - 9