Download Print this page

Interacoustics AD226 Instructions For Use Manual page 26

Diagnostic audiometer
Hide thumbs Also See for AD226:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 545
AD226 Instruction for Use - EN
Masking
Presentation
Intensity
Frequency range 125Hz to 8kHz.
Internal storage
Data Connections
(sockets) for connection
of accessories
External devices (USB)
Display
Compatible software
(optional)
Dimensions (LxWxH)
Weight
Power supply
Batteries
Operation environment
Transport and storage
Warm up time
Included Parts
Narrow band noise:
IEC 60645-1 2012, 5/12 Octave filter with the same centre frequency
resolution as pure Tone.
Synchronous masking: Locks channel 2 attenuator to channel 1 attenuator.
Manual or Reverse. Single pulse. Multiple pulses 50-5000 msec. on/off.
AC: -10 to 120 dB HL
BC: -10 to 80 dB
Available Intensity Steps is 1, 2 or 5dB
Extended range function: If not activated, the Air Conduction output will be
limited to 20 dB below maximum output.
Extended range only available when mains powered
125Hz, 250Hz, 500 Hz, 750Hz, 1500Hz or 8kHz may freely be deselected
500 patients
1 x USB A for keyboard or printer
1 x USB B for PC connection (compatible with USB 1.1 and later)
Standard PC keyboard (for data entry)
Supported printers: Please contact local distributer for a list of approved PC
printers.
4,3" (480x272) TFT color display.
Diagnostic Suite - Noah, OtoAccess and XML compatible
30x23x9cm, 12x9x4 inches.
1.3kg / 2.9lb
5VDC-max 1.6A UE10 type only
4x1.5V/1.2V Alkaline/NiMH Type AA,
Note:
When the instrument is battery operated the maximum stimuli output level
is reduced 20dB
Temperature:
Re. Humidity:
Ambient pressure:
Transport temperature:
Storage temperature:
Re. Humidity:
Approx. 1 minute
DD45 Audiometric headset
B71 Bone conductor
APS3 Patient response switch
Power cable
Operation manual CD
Multilingual CE Instructions for Use
15-35°C
30-90% Non condensing
98-104 kPa
-20-50°C
0-50°C
10-95% Non condensing
Page 18

Advertisement

loading