Addressing The Endurance Sensor In A Rs485 Multidrop Network - Fluke Endurance Series User Manual

Innovative high temperature fiber optic infrared pyrometers
Hide thumbs Also See for Endurance Series:
Table of Contents

Advertisement

multidrop manner, whereas all A-signals, as well all B-signals have to be connected to common
lines. The common A-signals have to be routed to the TX+ and the common B-signals to TX-
terminal at the selected USB/RS485 converter.
5.5.6. Addressing the Endurance
If you are installing two or more sensors in a multi-drop configuration, please be aware of the
following:
 Each sensor must have a unique address greater zero (1 - 32).
 Each sensor must be set to the same baud rate (default is 38.4 kBaud).
 Once all the units are addressed, wire up the units in the 2-wire multidrop manner,
keeping all A & B to be common.
 Now you can run the supplied Endurance
software or an individual terminal program to access the Endurance
commands and receive the responses.
Innovative High Temperature Fiber Optic Infrared Pyrometers
®
sensor in a RS485 Multidrop Network
®
software, an own written communication
®
Installation
sensor for issuing
33

Advertisement

Table of Contents
loading

Table of Contents