Fluke Endurance Series User Manual page 58

Innovative high temperature fiber optic infrared pyrometers
Hide thumbs Also See for Endurance Series:
Table of Contents

Advertisement

Endurance® Series
Users Manual
Order of subentry appearance:
1. Subentry: RS485 BAUD RATE
The subentry RS485 BAUD RATE is to set the RS485 communication baud rate,
whereat the default baud rate is set to 38400 bps
With the ▲ ▼ keys, you can toggle between the following communications baud rates:
1200 bps, 2400 bps, 9600 bps, 19200 bps, 38400 bps, 57600 bps, 115200 bps
2. Subentry: MULTIDROP ADDR.
The subentry MULTIDROP ADDR. is to assign a specific serial multidrop address to
an Endurance
environment, where several devices interact with each other.
With the ▲ ▼ keys, you can toggle between sub-addresses from 000 to 032
3. Subentry: TERMIN. RESISTOR
The subentry TERMIN. RESISTOR is to reduce signal reflections over long distance
connections by inserting a termination resistor of 120Ω.
With the ▲ ▼ keys, you can toggle between ON and OFF (120Ω insertion)
4. Subentry: ETHERNET DHCP
The subentry ETHERNET DHCP is to indicate to a network DHCP server, to obtain a
dynamic Ethernet address. The DHCP server assigns the Endurance
dynamic address out of an address pool.
With the ▲ ▼ keys, you can toggle between ON and OFF (dynamic address service)
5. Subentry: ETHERNET IP
The subentry ETHERNET IP is to set a fix unique network device address for the
®
Endurance
device, if DHCP in inactive. The assigned address has to fit in the network
address pool of your subnet.
The ENTER button selects in a consecutive way the IP address byte aaa.bbb.ccc.ddd
With the ▲ ▼ keys, you can toggle the individual address byte between 0 and 255
6. Subentry: ETHERNET NM
The subentry ETHERNET NM is to set a fix unique network mask address to integrate
the Endurance
in the network address pool of your subnet.
The ENTER button selects in a consecutive way the NM address byte aaa.bbb.ccc.ddd
With the ▲ ▼ keys, you can toggle the individual address byte between 0 and 255
7. Subentry: ETHERNET GW
The subentry ETHERNET GW is to set a fix unique network gateway address to
integrate the Endurance
has to fit in the network address pool of your subnet.
The ENTER button selects in a consecutive way the GW address byte aaa.bbb.ccc.ddd
With the ▲ ▼ keys, you can toggle the individual address byte between 0 and 255
8. Subentry: ETHERNET PORT
The subentry ETHERNET PORT is to set a fix port address for the relevant network
services of the Endurance
address is used for any special network request by the Endurance
With the ▲ ▼ keys, you can toggle the TCP/UDP port address from 0 to 65535
9. Subentry: WEB SERVER
The subentry WEB SERVER is to activate the Endurance
functionality for video and web based applications.
With the ▲ ▼ keys, you can toggle the WEB SERVER selector between OFF and ON
46
®
device, which is working in a 2-wire (half duplex) multidrop
®
device in an existing subnet domain. The assigned address has to fit
®
device in an existing subnet domain. The assigned address
®
device in an existing subnet domain. The assigned port
®
device a
®
device.
®
device internal web server