4.6. Sensor Placement (
The following figure demonstrates the sensor placement under various conditions, where valid
2-Color temperature measurements are possible. Note, however, that if the sensor signal is
reduced more than 95% (including emissivity and obscuration of the target), the sensor
accuracy also degrades.
Sighting hole smaller than the sensor's
Innovative High Temperature Fiber Optic Infrared Pyrometers
Mode)
2-Color
field of view
Dirty lens or dirty sighting window
Smoke, steam, dust, gas
in atmosphere

Figure 7: Sensor Placement in 2-Color Mode

Emitted
energy
Emitted
energy
Emitted
energy
Target smaller than field of view
and/or moves or vibrates in and out of
Emitted
field of view (e.g. wire)
energy
Sensor Location
21