Message process alarm
Behavior on module fault

8.3.4. Structure of the input/output data

8.3.4.1. Pyrometer module input data
The input data length is 23 Byte.
Address
Length
without offset
0
4 Byte
4
4 Byte
8
4 Byte
12
4 Byte
16
4 Byte
20
1 Byte
21
2 Byte
8.3.4.2. Pyrometer module output data
The output data length of Input/Output module is 5 Byte. The output data may be used to
change the initialization of the device (which was set once at start-up) when the bus is in data
exchange mode.
To do so the following structure is defined:
Address
Length
without offset
0
1 Byte
1
4 Byte
The <Type of parameter> gives the meaning of the following parameters (with the same format
as described in the section 8.3.3.1 for Pyrometer parameters)
Parameter type
0
Do not change anything
1
Slope
2
Emissivity
3
Transmissivity
4
Averaging time
5
Peak hold time
6
Valley hold time
7
Set point for the relay
8
Laser control
If <Type of parameter> is set to 0 then the output data gets ignored.
As default, it should be set to 0 (zero).

8.3.5. Diagnostics

The diagnostics information of the fieldbus communicator can be read out acyclically using
standard diagnostics data sets defined in the PROFINET IO specification.
Innovative High Temperature Fiber Optic Infrared Pyrometers
is transferred to the PROFINET
IO controller
The process alarm of pyrometer
is not transferred to the
PROFINET IO controller
is transferred to the PROFINET
IO controller
Format
REAL (Big Endian, Motorola)
REAL (Big Endian, Motorola)
REAL (Big Endian, Motorola)
REAL (Big Endian, Motorola)
DWORD
BYTE Bit0 (Bool)
INT(Big Endian, Motorola)
Format
BYTE
REAL/ WORD
(Big Endian, Motorola)
Meaning
Device Options
message active
message inactive
message active
set process data to zero,
set process data to last value
Value
Target Temperature 2 color
Target Temperature 1 color wide
Target Temperature 1 color narrow
Internal temperature
Error Code
Trigger state ( 0 – reset, 1 – set)
Measured attenuation
Value
Type of parameter
Parameter
Format
REAL
REAL
REAL
REAL
REAL
REAL
REAL
WORD
61