3051S SIS Specifications

Comprehensive Proof Test
The comprehensive proof test consists of performing the same steps as the
simple suggested proof test but with a two point calibration of the pressure
sensor in place of the reasonability check. This test will detect ~ 87% of
possible DU failures in the device.
Required tools: Field Communicator and pressure calibration equipment.
1. Bypass the safety function and take appropriate action to avoid a false
trip.
2. Perform Proof Test 1.
3. Perform a minimum two point sensor calibration check using the 4-20
mA range points as the calibration points.
4. Check the reference mA meter to verify the mA output corresponds to
the pressure input value.
5. If necessary, use "Choosing a Trim Procedure" on page 4-5 of the 3051S
Reference Manual.
6. Document the test results per your requirements.
7. Remove the bypass and otherwise restore normal operation.
8. Place the Security switch in the "ON" position.
NOTE
The user determines the proof test requirements for impulse piping.
Visual Inspection
Not required
Special Tools
Not required
Product Repair
The 3051S is repairable by major component replacement.
All failures detected by the transmitter diagnostics or by the proof-test must be
reported. Feedback can be submitted electronically at
http://rosemount.d1asia.ph/rosemount/safety/ReportAFailure_newweb.asp
All product repair and part replacement should be performed by qualified
personnel.
The 3051S must be operated in accordance to the functional and
performance specifications provided in the 3051S Product Data Sheet
(document number 00813-0100-4801).
Failure Rate Data
The FMEDA report includes failure rates and common cause Beta factor
estimates.
The report is available at
http://www2.emersonprocess.com/en-US/brands/rosemount/Safety-Products/Pages/index.aspx
Rosemount 3051S Series
.
7-39