Environmental - Fluke 93 Service Manual

Scopometer. philips pm93; pm95; pm97
Table of Contents

Advertisement

CHARACTERISTICS
2
-
19
CHARACTERISTICS
SPECIFICATIONS
ADDITIONAL INFORMATION
2.20
ENVIRONMENTAL
The
characteristics
are
valid
only
If
the instrument
is
checked
in
accordance
with the
official
checking
procedure.
Meets
Environmental
MIL-T-2a800D Type
111
Requirements
of:
Class
3,
Style
C
Temperature
Batteries
removed
from Instrument
Operating
0 "C...50
unless Pdttenes
meet
the required
tmperature
specifications.
Maximum
Operating
Temperature
derated 3
for
each
km. (each
3000
feet)
above sea
level.
Non
Operating (Storage)
20
^..70
"C
*
Maximum
Humidity
Non
Operating
(Storage)
95%
Relative
Humidity
Operating
20^C...30
90%
30
®C...50"C
70%
Maximum
Altitude
Memory
backup
batteries
removed
Operating
3
km(10
OOOfeet)
from
instrument unless
batteries
meet
maximum
altitude
specifications.
Non
Operating
(Storage)
12
km
(40
000
feet)
Vlpraoon
(Operating)
Frequency
5
w.
15
Hz
Sweep
Time 7
min.
Excursion
(pk
to
pk)
1.5
mm
Max
Acceleration
7 m/s*
(0.7
X
9)
@
15 Hz.
Frequency
15...
25
Hz
Sweep
Time
3
min.
Excursion
(pk
to
pk)
1.0
mm
Max
Acceleration
13 m/s*
(1.3
X
g)
@
25
Hz.
Frequency
25...
55
Hz
Sweep
Time
5
min.
Excursion
(pk
to
pk)
0.5
mm
Max
Acceleration
30 m/s*
(3.0
X
g)
@
55
Hz.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

9596