Fluke 93 Service Manual page 76

Scopometer. philips pm93; pm95; pm97
Table of Contents

Advertisement

PERFORMANCE
VERIFICATION
PROCEDURE
4-7
Proc^durd
for
channel
A
rice
time
meaaurament:
Refer
to
the settings/procedure
for
channel
B
measurement.
Requirements:
Refer
to
channel
B
requirements.
10/11/12/13.
Frequency response
These
tests
check
the
upper
transition
point of the
bandwidth
for
ScopeMeter
vertical
channels
Aand
B.
Test
equipment:
Tektronix
SG
503
Constant Amplitude Sine
wave
Generator
Test
setup:
Procedure/requirements
for
channel
A
frequency
response measurement:
A
Apply
a
50
kHz
sine
wave
with
an
amplitude
of
120
mV
peak-to-
peak
to
channel
A.
Use a
501^
termination.
Adjust the
input signal
to
a
trace height of exactly
6
divisions.
B
Without changing
the
amplitude
of
the
sine
wave
signal,
switch
over
to
step
1
1
using the
upper
select/adjust
key.
Increase
the
frequency
of
the
sine
wave
to
50
MHz
and
venfy
that
the
vertical
deflection
is
4.2
divisions
or
more.
Procedure/requiremente
for
channel
B
frequency response
measurement!
C
Apply a
5D
kHz
sine
wave
with
an amplitude
of
1
20
mV
peak-to-
peak
to
channel
B.
Use
a
SOiZ
termination.
Adjust
me
input signal
to
a
trace height of exactly
6
divisions.
0
Without changing
the
amplitude
of
the sine
wave
signal,
switch
over
to
step
13 using
the
upper
select/adjust
key.
Increase the frequency
of
the
sine
wave
to
50
MHz
and check
that
the
vertical
deflection
is
4.2
divisions or
more.