Fluke 93 Service Manual page 113

Scopometer. philips pm93; pm95; pm97
Table of Contents

Advertisement

CALIBRATION
ADJUSTMENT PROCEDURE
5-17
Tabi9
S.3
CaJibrat/on signals for
stop M4...M7.
Calibration
Step
Number
C^ibration Voltage
300
mV DC
3V
DC
30V
DC
M7
30OV
DC
WARNING:
After
you have performed
calibration
M7,
deactivate the
Fluke
5100B
to
remove
the
300V
DC. Always
set
the
Fluke
S1008
to
300
mV
DC
before touching the
connection
cables!
M8/9. External
input,
300
mV/3V
range: gain
calibration
Purpose:
calibration of
the
external
input
gain
in
the
300
mV
and
3V
ranges.
Calibration
equipment:
Fluke
51
006
Calibrator
Calibratjon setup:
Procedure:
A
Apply
300
mV DC
to
the
banana
connectors.
B
-
Press the
READY
softkey.
C
-
Apply
3V
DC
to
the
banana
connectors.
D
-
Press
the
READY
softkey.