4-4
PERFORMANCE
VERIFICATION
PROCEDURE
4.
Vertical
deflection coefficients
channei
A
These
tests
check
the
vertical deflection coefficients for
cttannel
A
in
the
100 mV/div
DC
and
AC
ranges.
Test
ec^uipment:
Fluke
5100B
Calibrator
Test setup:
Procedure/requlremente
for
AC
teat:
A
Apply
a
1
kHz
sine
wave
signal with
an
amplitude
of
600
mV AC
peak-to-peak
to
the
channel
A
BNC
connector.
(Set the Fluke
5100B
to
212.13
mV
RMS,
1
kHz
sine
wave).
Verify that
the amplitude
of
the sine
wave
signal
dieolayed
is
5.68...6.12
divisions.
Procedure/requirements
DC
test:
B
Apply
300
mV
DC
to
channel
A.
Verify that
the distance
between
the trace
for
channel
A
and
the
vertical
middle
of the
screen
(ground
level)
Id
2.94..
.3.06
divisions.
5/6/7.
Vertical
deflection coefficients ctiannel
D
These
tests
check
the
vertical
deflection coefficients
for
channel
6
In
the
DC
and
AC
ranges.
Test
equipment:
Fluke
5100B
Calibrator