7
-4fi
CORRECTIVE MAINTENANCE
7.
2.
5.6
Input
unit
assembly
replacement
Figure
7.
14 Irput
unit
assembly
Referring to
figure
7.14.
use
the
foilowing
procedure
for
Input
unH assembly
replaoOTeni.
1
.
Remove
the
front
cover assembly.
2
.
Disassem
bl
e
the
bottom cove
r
assemb
ly.
3.
Remove
the
30-pole
flat
cable.
4.
Unsolder
the wiring
(6x) of
the
input
terminals
from the analog
A2 PCB.
5-
The
input
unit
assembly
Is
clamped
onto
the
analog
A2
PCB. Loosen
these
clamps and
pull
the
input
unit
assembly
from
the
analog
A2
PCB.
NOTE:
The
white
gaskets
on
the input terminals
(4x)
are
sealed
to,
and must
remain
with
the input
unit
assembly.
Do
not
damage
the
gaskets
and do
not separate
them
from
the Input
unit
assembly.
Correctly
fitted
gaskets assure
the sealing of the
ScopeMeter.
6.
Reinstall
the
front unit
assembly.
7.
2. 5.
7
Top cover assembly
replacement.
Referring
to figure 7.12.
use
the following
procedure
for
top
cover
assembly
replacement.
1
.
Remove
i^e bottom cover assembly.
2.
Remove
the 30-pole
fiat
cable.
3.
The
digital
A1
PCB
and
metal screening are secured
to
the lop
cover
with four
M3
Torx
screws
(Item 33).
Two
of
these
screws
contain
standoffs,
be sure
to
put
them
on
the
right
place
again.
Use a
Torx screwdriver
to
remove
the screws.
4.
Remove
the metal
A1
screening from
digital
At
PCB.
5.
Lift
the
digital
A1
PCB
out
of
the top cover.
Be
careful not to
damage
the
infrared
LED
and
phototranslstor
of
the
optical Interlace.