6
PERFORMANJCE
VERIFICATION
PROCEDURE
Tost
setup channel
6
rise
Hme
measurement:
Proce<iiire
for
channel B
rise
time
measurement:
A
Apply a
fast rise
time
pulse, repetition
frequency
1
MHz,
amplitude
0.5V
to
channel
B.
Use
a
50Q
termination.
SetUie
generator
In
position
"FAST
RISE*.
B
Adjust the pulse am^^ltude
to
exactly
5
divisions.
See
figure 4.4.
Requirements:
A/OTE;
i^(
measured)
=
Inputsigriaf)
^
+
t^{
ScopeMeter)
®
C
Check
the
rise
time,
measured between
10%
and
90%
of the
pulse amplitude.
See
figure
4.4.
The
rise
time
t^
{measured)
must be 7
ns
(0.7
div)
or
less.
Figure 4.4
Rise
time
<o,7div
blbVm
Test setup
channel
A
rise
time
measurement:
Refer
to
the
test
set-up
for
channel B
measurement. Connect
the pulse generator
to
the channel
A
BNC
Input
connector.