5-14
CAUBRATION ADJUSTMENT
PROCEDURE
Calibration setup:
Procedure:
A
Apply a
1
MHz
square
wave
signal with
an amplitude
of
approximately
600
mV
peak-to-peak
to
both channel
A
and
B.
Setthe
generator
to the
FAST
RISE
position.
Use
50Q
terminations.
B
-
Press
the
READY
softkey.
C
Now
press
the
SCOPE
softkey
to
go back
to
tfie
CALIBRATE
menu.