Corrective Maintenance; Diagnostic Testing And Troubleshooting; Introduction; Troubleshooting Techniques - Fluke 93 Service Manual

Scopometer. philips pm93; pm95; pm97
Table of Contents

Advertisement

CORRECTIVE MAINTENANCE
7-1
7
CORRECTIVE MAINTENANCE
7.1
DIAGNOSTIC TESTING
AND TROUBLESHOOTING
7.1.1
Introduction
The
ScopeMeter
provides
semimodular
design
to
aid
in
troubleshooting.
This
section
describes
procedures
needed
to isolate
a
problem
In
a
specific functional
area.
Finally,
troubleshooting
hints for
each
functional
area are
presented.
If
the
ScopeMeter
fails,
first
verify that
you are
operating the
ScopeMeter
correctly
by
reviewing the
Operation
Venficatlon
Procedure
found
in
the
Users Manual.
WARNING:
Opening
the
case
may
expose hazardous
voltages.
Always
disconnect the
instalment
from
all
voltage
sources
and remove
the
batteries
before
opening
the
case.
Remember
that repairs
or
servicing
should be performed by
qualified
personnel
only.
7.1
.2
Troubleshooting techniques
If
a
fault
appears,
the following
test
sequence can be used
to
help
you
to
locate the defective
component:
Check
to
verify that
the
control settings
of
the
instrum^t
are
correct.
Consult
the operating
instructions
in
the
Users Manual.
Che<A
the
equipment
to
which
the
ir^strument
is
connected
and
check
the
interconnection cables.
Verify that
the instrument
is
properly
calibrated.
If
It
is
not,
refer to
Chapter
5:
'Calibration
Adjustment P^ocedure^
Locate the
circuit{6)
in
which
you
suspect the
fault:
the
symptom
often
suggests
the
faulty
circuit.
If
the
power
supply
is
defective,
the
symptom may
appear
to
be caused
by
several
circuts.
Check
the
drcult(s)
In
which you suspect
the
fauH.
Often
it
is
possible to
find faults
such as
cold
or defective solder
joints,
intermittent
or
open
interconnection plugs
and
wires or
damaged
components.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

9596

Table of Contents