Zeiss Humphrey II-i series User Manual page 127

Field analyzer
Table of Contents

Advertisement

3
Enter patient data. A patient name, date of birth, and Patient ID is
required for all of the following: saving to disk, STATPAC calculations,
automatic trial lens calculations, and screening tests using Age
Corrected mode.
4
fixation monitoring system to best suit your patient's needs.
5
patient test instructions. Adjust the table and perimeter to a
comfortable height for the patient. Make sure the patient is sitting
comfortably. Refer to
details of patient preparation. You also may refer to the Help Menus by
pressing the
the screen.
6
Press START.
7
If the Foveal Threshold parameter is turned on, it will activate now. Refer to the discussion of
"Foveal Threshold," on page
8
If the fixation monitoring parameter is set to "Gaze Track" or "Gaze/Blind Spot", follow the
on-screen instructions for setup. Refer to
page 5-4
for details. After gaze monitor initialization, a pop-up window appears and prompts
you to start the test. Remember to monitor the patient during the visual field test to insure
accurate results.
Humphrey Field Analyzer II-
i
series User Manual
Change test parameters. Select the test parameters and the
Patch the non-test eye. Reduce the room illumination. Give the
"Preparing the Patient," on page 3-24
HELP
icon (question mark icon) on the right-hand side of
5-7.
"Gaze Tracking (Models 740i, 745i, 750i)," on
Testing
5-17
for full