Attaching/Removing a Target Mask

The CM-700d/600d must be used with a Target Mask conforming to the selected lens position and
measurement condition.
To attach/remove a Target Mask, follow the procedure given below.
N
ote
• When attaching/removing a Target Mask, be careful not to allow dirt or dust to enter the integrating sphere
though the specimen measuring port.
• Do not touch the white-coated inner surface of the integrating sphere, wipe it with a cloth or put an object inside it.
• Do not exert excessive force on the latch of the Target Mask. Doing so may damage the latch, disabling use of
the Target Mask.
M
emo
If a Target Mask is damaged, contact a KONICA MINOLTA SENSING-authorized service facility.
Attachment/Removal to/from the Instrument
The figure shows the
Cut in the outer edge
• Attachment
1. Place the Target Mask over the specimen measuring
port so that the cut in the outer edge of the Target
Mask is aligned with the starting point of the
positioning mark (
2. Hold the outer edge of the mask, and turn it in the
direction of the arrow (clockwise). Turn the mask
until the cut in the outer edge is aligned with the "
mark on the instrument and then secure the mask.
• Removal
1. Hold the outer edge of the mask, and turn it in the
opposite direction of the arrow (counterclockwise).
Turn the mask until the cut in the outer edge is
aligned with the starting point of the positioning
mark (
) of the instrument.
2. Hold the outer edge of the mask and remove it.
Target Mask
φ
8 mm type (w/ plate) <For MAV>.
Plate
) on the instrument.
Target Mask positioning mark
Section of the CM-700d/
600d to which the Target
Mask is attached.
"