The easyTymp is intended for use in the electromagnetic environment specified below. The customer or the user
of the device is should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
level
Electrostatic
+6 kV contact
Discharge (ESD)
+8 kV air
IEC 61000-4-2
Electrical fast
+2 kV for power supply
transient/burst
lines
IEC 61000-4-4
+1 kV for input/output
lines
Surge
+1 kV differential mode
+2 kV common mode
IEC 61000-4-5
Voltage dips, short
< 5 % UT
interruptions and
(>95 % dip in UT) for 0.5
voltage variations on
cycle
power supply lines
40 % UT
IEC 61000-4-11
(60 % dip in UT) for 5
cycles
70% UT
(30% dip in UT) for 25
cycles
<5 % UT
(>95 % dip in UT) for 5
sec
Power frequency
3 A/m
(50/60 Hz)
IEC 61000-4-8
Note: UT is the A.C. mains voltage prior to application of the test level.
8108757 Rev. 3
Operation Manual easyTymp
Compliance
+6 kV contact
+8 kV air
+2 kV for power supply
lines
+1 kV for input/output
lines
+1 kV differential mode
+2 kV common mode
< 5 % UT (>95 % dip in
UT)
for 0.5 cycle
40 % UT (60 % dip in
UT) for
5 cycles
70% UT (30 % dip in
UT) for
25 cycles
<5 % UT
3 A/m
53
Electromagnetic
Environment-Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be greater than
30%.
Mains power quality should be that
of a typical commercial or
residential environment.
Mains power quality should be that
of a typical commercial or
residential environment.
Mains power quality should be that
of a typical commercial or
residential environment. In case of
a power failure the device will
automatically shut down within 10
s. If the user of the device requires
continued operation during power
mains interruptions, it is
recommended that the device be
powered from an uninterruptable
power supply or its battery.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or residential
environment.
15/11/2016