Maico easyTymp Operation Manual page 8

Hide thumbs Also See for easyTymp:
Table of Contents

Advertisement

Operation Manual easyTymp
1.7.3 Acoustic Reflex
An Acoustic Reflex, or contraction of the stapedial muscle, occurs under normal
conditions when a sufficiently intense sound is presented to the auditory pathway. This
contraction of the muscle causes a stiffening of the ossicular chain which changes the
compliance of the middle ear system. As in Tympanometry, a probe tone is used to
measure this change in compliance.
When the stimulus presentation and measurement are made in the same ear by means
of the probe, this acoustical reflex is referred to as an Ipsilateral Acoustic Reflex. When
the stimulus presentation is made in the opposite ear of where the measurement is made,
this acoustical reflex is referred to as a Contralateral Acoustic Reflex.
For best results, this reflex measurement is automatically conducted at the air pressure
value where the compliance peak occurred during the Tympanometric test. Stimulus
tones of varying intensities at 500 Hz, 1000 Hz, 2000 Hz or 4000 Hz are presented as
short bursts. If a change in compliance greater than the selected value is detected, a
reflex is considered present. Because this is an extremely small compliance change, any
movement of the probe during the test may produce an artifact (false response). The test
result is recorded as Pass/No Response, and in graphical form.
If the Tympanometric results display any abnormal findings, the results of the Acoustic
Reflex testing may be inconclusive and should be interpreted with care. Theoretically, a
compliance peak is necessary to observe a reflex at peak pressure.
1.7.4 Contralateral Acoustic Reflex
A Contralateral Acoustic Reflex is available with the easyTymp Pro Version. When the
stimulus presentation and measurement are made in the different ears by means of the
Contra Probe.
1.7.5 Acoustic Reflex Decay
An Acoustic Reflex Decay is available with the easyTymp Pro. Acoustic reflex decay, also
known as adaptation, is the measurement of the Acoustic reflex response during
sustained stimulus presentation. Ipsilateral and Contralateral Reflex Decay can be
performed.
1.7.6 Eustachian Tube Function Test
The Eustachian tube connects the middle ear with the nasopharynx. Its function is to
equalize pressure between the middle ear and the atmosphere.
The Eustachian tube test can be used to determine if the Eustachian tube is functioning
properly in patients with an intact tympanic membrane or in patients who have a
perforated TM or pressure equalization tubes.
15/11/2016
8108757 Rev. 3
5

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the easyTymp and is the answer not in the manual?

Table of Contents

Save PDF