Maico easyTymp Operation Manual page 21

Hide thumbs Also See for easyTymp:
Table of Contents

Advertisement

ISSUE
SOLUTION
Extension
If the device shows leaking, please
cable
1. Follow the suggestions for probe tip/ Probe.
2. If step 1 is not helpful, please change the extension cable. If the
Battery slot
1. If the spare battery is not charging, please, check if the battery is
2. Please make sure the battery contacts are clean in the case.
Connection
1. Make sure the handheld is properly inserted after the test.
in cradle
2. Please make sure battery contacts are clean in the case.
Printer
1. Please check if the cradle is connected with power supply.
problem
2. Please check if the printer function is activated in the device.
3. Please check if the printer paper is properly inserted.
4. Properly place the handheld on the cradle.
5. When during printing process the 2
PC
1. Make sure the Patient database and the printer is deactivated
Connections
2. Handheld:
3. Cradle:
4. Make sure the easyTymp option is selected in the PC software
5. Try to reinstall the PC software. Check the device manager in
8108757 Rev. 3
problem persists follow the suggestions for Probe tip/Probe.
properly inserted and the terminals are in contact (springs in
cradle).
Improper docking may lead to no connection between device
and the cradle.
nd
take the 2
battery out of cradle and try again.
from handheld.
a. Please check the USB connection in the PC and the
system.
b. Use another USB cable.
a. Make sure the device is properly placed into the Cradle.
b. Make sure the Cradle is powered while transferring the
result to PC.
(for detail contact your distributor).
the PC. If the easyTymp does not appear in the list install the
driver again using the installation CD.
18
Operation Manual easyTymp
nd
battery is charged also,
15/11/2016