1.5 Essential Performance
The following is considered essential performance:
To generate and present stimulus signals in the audio and pressure ranges as specified
in the applicable IEC 60645/ANSI S3.39 series in normal condition.
It is intended to be used by audiologists, ENTs, hearing healthcare professionals, or other
trained technicians in a hospital, clinic, healthcare facility or other suitable quiet
environment as defined in standard ISO 8253-1.
The easyTymp can be used on patients in the age range of infants, children and adults.
1.6 Features and Benefits of the easyTymp
The purpose of the easyTymp test system is to provide a rapid Tympanometry and
Acoustic reflex measurements to measure the middle ear status where a pass or refer
notation is identified. easyTymp provides an optional 1 kHz probe tone for testing newborns.
Factory defined protocols allow for simple screening measurements, and different versions
are available that provide diagnostic testing functions. As with any type of hearing
screening, a "pass" result should not overrule any additional concerns regarding middle
ear function. A referral to physician should be administered if concerns about middle ear
function persists.
The easyTymp cradle serves as a docking and recharging station for the handheld device
and can include an integrated printer or opening for placement of the eartip box.
Using the included Software, the handheld unit will transfer data to a PC via USB-
connection while in the docking station, or it can also transfer data directly via USB cable
when no docking station is available.
The easyTymp comes in multiple versions and configurations dependent on the country
and service partner. Each version provides specific testing functionalities dependent upon
the user needs.
easyTymp (as Standard Version)
Rapid Tympanometry measurement
Ipsilateral Acoustic Reflexes at several frequencies
1 kHz probe tone (option)
easyTymp Pro Version (Contra probe required)
Rapid Tympanometrymeasurement
Ipsilateral Acoustic Reflexes at several frequencies
Contralateral Acoustic reflex measurements at several frequencies
Acoustic Reflex Decay (Ipsilateral and Contralateral)
Eustachian Tube Function
1 kHz probe tone (option)
8108757 Rev. 3
Operation Manual easyTymp
3
15/11/2016