Cooling Parameters; Switch-On Delay Of The Standby Cooling - Hettich UNIVERSAL 320 R Repair Instructions

Hide thumbs Also See for UNIVERSAL 320 R:
Table of Contents

Advertisement

6.12

Cooling parameters

6.12.1

Switch-on delay of the Standby cooling

After a centrifugation run the standby cooling takes place time-delayed.
The delay time is preset to 1 minute.
The delay time can only be set, if the rotor is at standstill and the lid is opened.
It is at any time possible to terminate the procedure by pressing the
STOP / OPEN
If no key is pressed for a period of 8 seconds the menu will be terminated
automatically.
Action
1. Switch on the mains
switch
2. Keep the
key
pressed for 8 seconds
Display after
8 seconds
• Turn the control
knob
3. Press the
key
START / IMPULS
key.
Display / Comment
UNI 320R Vx.xx
Machine type and program version
# 30 4000 01:00
Display of the centrifugation data
# 30 4000 01:00
t/min
=
Delay time of the standby cooling
→ t/min
=
Set the delay time.
The delay time is settable from 1 to 5 minutes, in
1 minute increments.
→ *** ok ***
The setting will be stored.
The menu will be terminated automatically.
X
X
41/66

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Universal 320

Table of Contents