6.4
Blank the machine type in the display
The display of the machine type after switching on the unit can be blanked.
It is at any time possible to terminate the procedure by pressing the
STOP / OPEN
Action
1. Switch on the mains
switch
2. While UNI 320R Vx.xx is
displayed, keep the
key and
START / IMPULS
the
key
STOP / OPEN
pressed simultaneously
for 16 seconds
3. Press the
key so
SELECT
often until the following
will be displayed
4. Press the
START / IMPULS
key
5. Press the
key so
SELECT
often until the following
will be displayed
• Turn the control
knob
6. Press the
START / IMPULS
key
26/66
key twice.
Display / Comment
→ UNI 320R Vx.xx
Machine type and program version
→ * SELECT MODE *
If KEYBOARD-ERROR will be displayed, a key
was briefly released.
The entire procedure must be carried out again.
→ GOTO SETUP
→ * SETUP-MODE *
→ OEM VERSION OFF
→ OEM VERSION ON
ON = machine type will not be displayed
OFF = machine type will be displayed
→
*** ok ***
The setting will be stored
→
*** exit ***
The menu will be terminated automatically