States Per Division Field (State only)

When the analyzer Type field in the Configuration menu is set to State, the
analyzer uses external clocks from the system under test. In this mode, the
X axis of the waveform display is measured in states per division.
You use the states/Div field to select the states per division resolution of the
X axis. You can specify between 1 and 500 states per division by touching
the states/Div field and rotating the knob, or by touching the states/Div field
twice and then using the pop-up keypad. By adjusting the states/Div, you can
zoom in to view a desired part of the display.
States per division
field
States Per Division Field
The Waveform Menu