Instrument diagnosis

Important operating values and instrument data are displayed. A
gas path check (testo 330-2 LL) can be carried out. The status of
the sensors and any device errors not yet rectified can be
displayed.
Calling up the function:
>
[
]
Instrument Diagnosis
or
>
[ i
].
Carrying out a gas path check (testo 330-2 LL)
1.
Gas Path Check
2. Place the black sealing cap on the tip of the flue gas probe.
-
The pump flow is displayed. If the volumetric flow rate is less
than 0.02 l/min, the gas paths are not leaking.
3. End of check: [OK].
Viewing device errors:
>
Error Diagnosis
-
Unrectified errors are displayed.
> View next / previous error: [▲], [▼].