CP03x/CP03xA Current Probes

Test Equipment Required

The following table lists the test equipment and accessories (or their equivalents)
which are required for performance verification of the CP03x/CP03xA Current Probes.
Because the input and output connector types may vary on different brands and
models of test instruments, additional adapters or cables may be required.
Description
Wide Band Oscilloscope
Digital Multimeter (2)
Function Generator
Calibration Fixture
Calibration Fixture
Banana Plug Adapter
Patch Cables (4)
BNC Adapter

Preliminary Procedure

1. Connect the probe to the C1 input of the instrument and completely close the
probe slider.
2. Turn on the instrument and allow at least 30 minutes warmup time for the probe
before performing the verification procedure.
3. Turn on the other test equipment and allow these to warm up for the time
recommended by the manufacturer.
4. While the instruments are reaching operating temperature, make a photocopy of
the Performance Verification Test Record and fill in the necessary data.
14
Minimum Requirements
200 MHz bandwidth
ProBus interface equipped
DC: 0.1% Accuracy
5½ digit resolution
50 Hz sine wave output 3 Vrms
into 50Ω
100 turn loop in series with 0.5Ω
±0.1% resistor with sense
terminals
ProBus Extension Cable
Female BNC to Dual Banana Plug
Male Banana to Male Banana, 12"
BNC Male to Dual Banana Jack
Example Equipment
Teledyne LeCroy WaveRunner 6 Zi
Agilent Technologies 34401A
Fluke 8842A-09
Teledyne LeCroy WaveStation 3082
Stanford Research Model DS340
Teledyne LeCroy CP031-CF02
Teledyne LeCroy PROBUS-CF01
Pomona 1269
Pomona B-12-0 (black)
Pomona B-12-2 (red)
Pomona 1296