Care And Maintenance; Maintenance - Beckman Coulter Avanti J-26S XP Series Instructions For Use Manual

Hide thumbs Also See for Avanti J-26S XP Series:
Table of Contents

Advertisement

JS-5.3 Swinging-Bucket Rotor

Care and Maintenance

8000
7000
6000
5000
4000
3000
2000
1000
0
0
Care and Maintenance

Maintenance

Do not use sharp tools on the rotor that could cause scratches in the rotor surface. Corrosion begins in
scratches and may open fissures in the rotor with continued use.
1
Periodically (at least monthly) inspect the rotor yoke and buckets, especially inside cavities, for
rough spots or pitting, cracks, white powder deposits—frequently aluminum oxide—or heavy
discoloration.
If any of these signs are evident, do not run the rotor.
Contact your Beckman Coulter representative for information about the Field Rotor Inspection
Program and the rotor repair center.
2
Before using the rotor, inspect the rotor drive pins to ensure that they are not damaged.
Damaged drive pins can prevent the rotor from seating properly on the centrifuge drive hub
(Avanti J series centrifuges only).
To inspect the drive pins, turn the rotor upside down and look into the drive hole in the center
of the rotor.
If the drive pins appear damaged, contact Beckman Coulter Field Service.
3
Before using the rotor tie-down knob (368410), check it for damage such as distortion, splitting,
or stripped threads.
Replace it if it is damaged.
1-16
Relative Centrifugal Fields for the JS-5.3 Rotor with 500-mL Conical Bottles
1000
2000
3000
4000
Speed (RPM)
r max
r av
r min
5000
5300
PN J-TB-089AG

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Avanti j-26 xp seriesAvanti jxn-26Avanti j-e