TROUBLESHOOTING

9.3 TROUBLESHOOTING GUIDE

Problem
DL.CODE Installation:
Autorun or Start.hta doesn't
run
Power ON:
the "POWER" LED is not lit.
One Shot or Phase Mode
using the Input 1 (External
Trigger) or Input 2:
the "TRIGGER" LED is not
blinking while the External
Trigger is switching.
One Shot or Phase Mode
using serial trigger source:
the "TRIGGER" LED is not
blinking.
Phase Mode:
the
"TRIGGER"
correctly blinking but no image
is displayed in the DL.CODE
window.
Continuous Mode:
the "TRIGGER" LED is not
blinking.
TROUBLESHOOTING GUIDE
Suggestion
Check Windows settings to see if Autorun is disabled.
Associate the file type .hta with the Microsoft HTML
Application host mshta.exe in Windows\System32.
Is power connected?
If using a power adapter (like PG6000), is it connected
to wall outlet?
If using rail power, does rail have power?
If using CBX, does it have power (check switch and
LED)?
Check if you are referring to the device / accessory
cable connector or to the CBX spring clamp connectors.
Measure Voltage either at pin 1 and pin 2 (for 17-pin
connector) or at spring clamp Vdc and GND (for CBX).
Check if you are referring to the device / accessory
cable connector or to the CBX spring clamp connectors.
Is the sensor connected to Input 1 or Input 2?
Is power supplied to the photo sensor?
For NPN configuration, is power supplied to one of the
two I1 or I2 signals (A or B)?
For PNP configuration, is one of the two I1 or I2 signals
grounded (A or B)?
Are the photo sensor LEDS (if any) working correctly?
Is the sensor/reflector system aligned (if present)?
On the Reading Phase step check the Input 1 or Input 2
Debouncing Time parameter setting.
On the Reading Phase step check the settings for
Acquisition
Reading Phase-OFF parameters.
On the Reading Phase step check the settings for
Acquisition
Reading Phase-OFF parameters.
Are the COM port parameters (Baud Rate, Parity, Data
Bits, Stop Bits) correctly assigned?
On the Reading Phase step check the settings of
Acquisition Trigger String, Reading Phase-ON
String, and Reading Phase-OFF String parameters.
Is the serial trigger source correctly connected?
Is the Phase frequency lower than the maximum frame
LED
is
rate?
Verify the correct software configuration settings.
Trigger,
Reading
Trigger,
Reading
9
Phase-ON,
and
Phase-ON,
and
117