1. DEFINITION
The "Bar code" option enables to install a bar code reader on a second specific RS232
connector of the ATEQ instruments.
It enables on reading of the code to select a test program and eventually to launch the test of
control (if the option is validated).
The quantity of characters read by the reader should not exceed 22. Beyond the unit will not
take into account the character string (bar code string).
1.1. C
OMMUNICATION PARAMETERS CONFIGURATION
1.1.1. Bar code reader configuration
The Bar Code reader must be set with the following communication parameters:
 9600 bauds,
 7 bits,
 1 stop,
 Parity even.
The communication parameters of the ATEQ devices are the same by default and fixed, it's not
provide to change it.
Version 1.04a
BAR CODE (OPTION)
Sheet # 694u – Bar code (option)
User guide ATEQ 6th series Page 1/10

Advertisement

loading

Summary of Contents for Datalogic Gryphon GD4100

  • Page 1 Sheet # 694u – Bar code (option) BAR CODE (OPTION) 1. DEFINITION The “Bar code” option enables to install a bar code reader on a second specific RS232 connector of the ATEQ instruments. It enables on reading of the code to select a test program and eventually to launch the test of control (if the option is validated).
  • Page 2 Sheet # 694u – Bar code (option) 1.1.2. RS232 port configuration From the "CONFIGURATION" menu, select the "AUTOMATISM" menu and then press the button By using the arrows, select the "RS232" menu and validate with the key. By using the arrows, select the "CODE READER"...
  • Page 3 DATALOGIC Gryphon I GD4100 (USB). For the reader configuration, follow this process: 1) Enter in programming mode of the reader by flashing this code (right hand) "ENTER/EXIT PROGRAMMING MODE".. 2) Reset the reader to factory settings by flashing this code (right hand) "Factory Default Settings".
  • Page 4: Function Activation

    Sheet # 694u – Bar code (option) 3. FUNCTION ACTIVATION Warning : at each parameter or configuration change will delete all the learnings. That will necessary to learn all the codes for each program. From the "CONFIGURATION" menu, select the "AUTOMATISM" menu and then press the button By using the arrows, select...
  • Page 5 Sheet # 694u – Bar code (option) By using the arrows, select the "Pr. Select" menu select by "Yes" with the arrows and validate with the key. The 'First Char." and "Char Number" are displayed, By using the arrows, select "First Char" and validate with key.
  • Page 6 Sheet # 694u – Bar code (option) "RESET EOC" parameter. Adjusted on "No" the device will save in memory the readed bar code for all the following programs, until a new bar code will be readed. Adjusted on "Yes" it will must flash a new bar code before each start cycle.
  • Page 7: Function Configuration

    Sheet # 694u – Bar code (option) 4. FUNCTION CONFIGURATION Activate the function or check if it checked. Press the key, the cursor slide to the right hand. By using the arrows, select "Yes" and validate with the key. The configuration menu appears. The "NUM."...
  • Page 8 Sheet # 694u – Bar code (option) 5. PARAMETER SETTING OF THE STRINGS (LEARNING) The characters string learning will be done from the special cycles. From the main menu, enter the special cycle menu. In the special cycle menu, select the special cycle "CODE READER", and validate with The device ask for the corresponding test program to the characters strings which be...
  • Page 9 Sheet # 694u – Bar code (option) Then flash the code by using the bar code reader. The captured characters are displayed. The code is recorded, the device is ready to run. At each reading of this code, the device will select the corresponding program.
  • Page 10 Sheet # 694u – Bar code (option) 6. FRAMES Examples of frames sended by the device, with the program bar code. 6.1. S TANDARD MODE FRAMES <01>:  <01>:13/10/2014 20:10:46  <01>: 0.598 bar:(OK):  78 Pa  <01>:123456789001XXXXXXXX01    <02>:  <02>:13/10/2014 20:10:57  <02>: 0.597 bar:(OK):  85 Pa  <02>:1234567890023333333301    <12>:  <12>:13/10/2014 20:11:06  <12>: 0.597 bar:(OK):  85 Pa  <12>:1234567890124444444412  6.2.