Residual Spurious Response - Anritsu S412E Maintenance Manual

Lmr master vector network/spectrum analyzer
Hide thumbs Also See for S412E:
Table of Contents

Advertisement

3-3 Residual Spurious Response

3-3
Residual Spurious Response
The following test is used to verify the residual spurious response of the spectrum analyzer of the S412E.
Equipment Required
• 50 Ohm Termination, Anritsu Model 28N50-2
Procedure
1. Connect the 50 Ohm Termination to the RF In connector of the S412E.
2. Power on the S412E and put the instrument into Spectrum Analyzer mode with the Menu button.
3. Preset the S412E by pressing the Shift key, then the Preset (1) key, then the Preset button on the
touchscreen.
Note
Before continuing, allow a 30-minute warm up for the internal circuitry to stabilize.
4. Press the Amplitude submenu button.
5. Touch the Detection submenu button and then the Peak submenu button.
6. Press the Sweep menu, then change the Sweep Mode to Performance.
Residual Spurious Test with Preamp Off
7. Set the Reference Level to –40 dBm using the Amplitude menu on the touchscreen.
8. Set the Attenuation Level to 0 dB using the Atten Lvl submenu.
9. Make sure that the Pre Amp On/Off submenu button is in the Off position. If the preamp is on, then touch
the Pre Amp On/Off submenu button to turn it off.
10. Set the Start Freq to 10 MHz using Freq menu on the touchscreen.
11. Set the Stop Freq to 1.6 MHz.
12. Set the RBW to 3 kHz using the BW menu.
13. Set the VBW to 1 kHz.
14. Wait until one sweep is completed.
15. Touch the Marker main menu button and then the Peak Search submenu button.
16. Verify that Marker 1 amplitude is < –90 dBm.
If a spur larger than –90 dBm appears, wait another full sweep and observe whether the spur
Note
re-appears at the same point on the second sweep. If the spur does not appear at the same point on
the second sweep, then the spur on the first sweep was not real.
17. Record the Marker 1 amplitude reading in
with Preamp Off" on page
18. Repeat this procedure with a Start Frequency of 1.6 GHz and Stop Frequency of 6 GHz if Option 6 is
installed, and record results in
3-4
Table A-3, "Spectrum Analyzer Residual Spurious Response
A-3.
Table
A-3.
PN: 10580-00324 Rev. N
Chapter 3 — Spectrum Analyzer Verification
S412E MM